A comparison between quadrupole, double focusing and multicollector ICP-MS instruments : Part I. Evaluation of total combined uncertainty for lead isotope ratio measurements
- 1 March 2001
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Journal of Analytical Atomic Spectrometry
- Vol. 16 (4), 315-321
- https://doi.org/10.1039/b006145j
Abstract
A quadrupole-based instrument (HP-4500), a double focusing single collector instrument (Element) and a single focusing multicollector (IsoProbe) were all evaluated for the precise and accurate measurement of lead isotope ratios using Tl as internal isotope ratio standard. The isotope ratio accuracy was evaluated using NIST 981 Common Lead Isotope standard reference material and an enriched 204Pb spike. Mass bias correction was performed using an exponential model in all the three instruments and equations for the evaluation of the total combined uncertainty arising from the correction for the dead time and mass bias were developed. It was observed that, in the correction for mass bias, the uncertainty in the natural Tl ratio was the main source of uncertainty for the multicollector instrument while, for the quadrupole and double focusing single collector instruments, the measured isotope ratio was the main source of uncertainty.Keywords
This publication has 35 references indexed in Scilit:
- Examination of the performance exhibited by a single detector double focusing magnetic sector ICP-MS instrument for uranium isotope abundance ratio measurements over almost three orders of magnitude and down to pg g−1 concentration levelsJournal of Analytical Atomic Spectrometry, 2000
- Determination of cadmium in environmental and biological reference materials using isotope dilution analysis with a double focusing ICP-MS: a comparison with quadrupole ICP-MSJournal of Analytical Atomic Spectrometry, 1999
- Accurate determination of iron, copper and zinc in human serum by isotope dilution analysis using double focusing ICP-MSJournal of Analytical Atomic Spectrometry, 1999
- A calculation method based on isotope ratios for the determination of dead time and its uncertainty in ICP-MS and application of the method to investigating some features of a continuous dynode multiplierJournal of Analytical Atomic Spectrometry, 1999
- Precision and accuracy in isotope ratio measurements by plasma source mass spectrometryJournal of Analytical Atomic Spectrometry, 1998
- Lead from Carthaginian and Roman Spanish Mines Isotopically Identified in Greenland Ice Dated from 600 B.C. to 300 A.D.Environmental Science & Technology, 1997
- Lead isotopic analyses of NIST Standard Reference Materials using multiple collector inductively coupled plasma mass spectrometry coupled with a modified external correction method for mass discrimination effectThe Analyst, 1996
- Determination of fission products and actinides by inductively coupled plasma-mass spectrometry using isotope dilution analysis: A study of random and systematic errorsAnalytica Chimica Acta, 1995
- Determination of lead and lead isotope ratios in gasoline by inductively coupled plasma mass spectrometryJournal of Analytical Atomic Spectrometry, 1994
- Effective deadtime of pulse-counting detector systemsAnalytical Chemistry, 1978