The influence of different technique factors on image quality of chest radiographs as evaluated by modified CEC image quality criteria
- 1 January 2002
- journal article
- research article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 75 (889), 38-49
- https://doi.org/10.1259/bjr.75.889.750038
Abstract
The Commission of the European Communities (CEC) research project "Predictivity and optimisation in medical radiation protection" addressed fundamental operational limitations in existing radiation protection mechanisms. The first part of the project aimed at investigating (1) whether the CEC image quality criteria could be used for optimization of a radiographic process and (2) whether significant differences in image quality based on these criteria could be detected in a controlled project with well known physical and technical parameters. In the present study, chest radiographs on film were produced using healthy volunteers. Four physical/technical parameters were varied in a carefully controlled manner: tube voltage (102 kVp and 141 kVp), nominal speed class (160 and 320), maximum film density (1.3 and 1.8) and method of scatter reduction (grid (R=12) and air gap). The air kerma at the entrance surface was measured for all patients and the risk-related dose H(Golem), based on calculated organ-equivalent dose conversion coefficients and the measured entrance air kerma values, was calculated. Image quality was evaluated by a group of European expert radiologists using a modified version of the CEC quality criteria. For the two density levels, density level 1.8 was significantly better than 1.3 but at the cost of a higher patient radiation exposure. The correlation between the number of fulfilled quality criteria and H(Golem) was generally poor. An air gap technique resulted in lower doses than scatter reduction with a grid but provided comparable image quality. The criteria can be used to highlight optimum radiographic technique in terms of image quality and patient dose, although not unambiguously. A recommendation for good radiographic technique based on a compromise between image quality and risk-related radiation dose to the patient is to use 141 kVp, an air gap, a screen-film system with speed 320 and an optical density of 1.8.Keywords
This publication has 6 references indexed in Scilit:
- The influence of different technique factors on image quality of lumbar spine radiographs as evaluated by established CEC image criteria.The British Journal of Radiology, 2000
- Calculation of Patient Doses Using a Human Voxel Phantom of Variable DiameterRadiation Protection Dosimetry, 2000
- Evaluation of lumbar spine images with added pathologyPublished by SPIE-Intl Soc Optical Eng ,2000
- Comparison of two methods for evaluating image quality of chest radiographsPublished by SPIE-Intl Soc Optical Eng ,2000
- In search of optimum chest radiography techniquesThe British Journal of Radiology, 1993
- Simultaneous Statistical InferencePublished by Springer Nature ,1981