Transmembrane Voltage Changes Produced by Real and Virtual Electrodes During Monophasic Defibrillation Shock Delivered by an Implantable Electrode
- 1 September 1997
- journal article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 8 (9), 1031-1045
- https://doi.org/10.1111/j.1540-8167.1997.tb00627.x
Abstract
Epicardial point stimulation produces nonuniform changes in the transmembrane voltage of surrounding cells with simultaneous occurrence of areas of transient positive and negative polarization. This is the phenomenon of virtual electrode. We sought to characterize the responses of epicardial ventricular tissue to the application of monophasic electric shocks from an internal transvenous implantable cardioverter defibrillator (ICD) lead. Langendorff-perfused rabbit hearts (n = 12) were stained with di-4-ANEPPS. A 9-mm-long distal electrode was placed in the right ventricle. A 6-cm proximal electrode was positioned horizontally 3 cm posteriorly and 1 cm superiorly with respect to the heart. Monophasic anodal and cathodal pulses were produced by discharging a 150-microF capacitor. Shocks were applied either during the plateau phase of an action potential (AP) or during ventricular fibrillation. Leading-edge voltage of the pulse was 50 to 150 V, and the pulse duration was 10 msec. Transmembrane voltage was optically recorded during application of the shock, simultaneously from 256 sites on a 11 x 11 mm area of the anterior right ventricular epicardium directly transmural to the distal electrode. The shock effect was evaluated by determining the difference between the AP affected by the shock and the normal AP. During cathodal stimulation an area of depolarization near the electrode was observed, surrounded by areas of hyperpolarization. The amplitude of polarization gradually decreased in areas far from the electrode. Inverting shock polarity reversed this effect. ICD monophasic defibrillation shocks create large dynamically interacting areas of both negative and positive polarization.Keywords
This publication has 20 references indexed in Scilit:
- Shock-Induced Depolarization of Refractory Myocardium Prevents Wave-Front Propagation in DefibrillationCirculation Research, 1996
- Optical Transmembrane Potential Measurements During Defibrillation-Strength Shocks in Perfused Rabbit HeartsCirculation Research, 1995
- Optical mapping of repolarization and refractoriness from intact hearts.Circulation, 1994
- Electrical stimulation of cardiac tissue: a bidomain model with active membrane propertiesIEEE Transactions on Biomedical Engineering, 1994
- Epicardial mapping of ventricular defibrillation with monophasic and biphasic shocks in dogs.Circulation Research, 1993
- Synchronized repolarization after defibrillation shocks. A possible component of the defibrillation process demonstrated by optical recordings in rabbit heart.Circulation, 1992
- Virtual cathode effects during stimulation of cardiac muscle. Two-dimensional in vivo experiments.Circulation Research, 1991
- Mechanism of cardiac defibrillation in open-chest dogs with unipolar DC-coupled simultaneous activation and shock potential recordings.Circulation, 1990
- Comparison of the defibrillation threshold and the upper limit of ventricular vulnerability.Circulation, 1986