Measurements of Particle Loss Functions in a Differential Mobility Analyzer (TSI, Model 3071) for Different Flow Rates
Open Access
- 1 January 1986
- journal article
- research article
- Published by Taylor & Francis in Aerosol Science and Technology
- Vol. 5 (4), 483-486
- https://doi.org/10.1080/02786828608959112
Abstract
Particle losses in a differential mobility analyzer (TSI, Model 3071) caused by diffusive deposition and Brownian diffusion are measured for particles in the diameter size range between 3 and 100 nm. For small sampling and aerosol flow rates (0.3 liters/min) at 20 nm, nearly 50% of the primary particles are lost; and for 2 liters/min, the particle losses have to be considered in the diameter size range below 30 nm (50% at 7 nm). From the measured penetration values, an effective tube length is derived which is useful to calculate particle losses for other flow rates through the analyzer.Keywords
This publication has 8 references indexed in Scilit:
- Bipolar diffusion charging of aerosol particles—II. Influence of the concentration ratio of positive and negative ions on the charge distributionJournal of Aerosol Science, 1984
- Bipolar diffusion charging of aerosol particles—I: experimental results within the diameter range 4–30 nmJournal of Aerosol Science, 1983
- Generation of monodisperse Ag- and NaCl-aerosols with particle diameters between 2 and 300 nmJournal of Aerosol Science, 1983
- Analysis of diffusion battery dataJournal of Aerosol Science, 1979
- Accurate measurement of aerosol electric mobility momentsJournal of Aerosol Science, 1975
- Aerosol classification by electric mobility: apparatus, theory, and applicationsJournal of Aerosol Science, 1975
- A submicron aerosol standard and the primary, absolute calibration of the condensation nuclei counterJournal of Colloid and Interface Science, 1974
- Electric aerosol particle counting and size distribution measuring system for the 0.015 to 1 μ size rangeTellus, 1966