Intensity modulated proton therapy: A clinical example
Top Cited Papers
- 9 March 2001
- journal article
- Published by Wiley in Medical Physics
- Vol. 28 (3), 317-324
- https://doi.org/10.1118/1.1350587
Abstract
In this paper, we report on the clinical application of fully automated three-dimensional intensity modulated proton therapy, as applied to a 34-year-old patient presenting with a thoracic chordoma. Due to the anatomically challenging position of the lesion, a three-field technique was adopted in which fields incident through the lungs and heart, as well as beams directed directly at the spinal cord, could be avoided. A homogeneous target dose and sparing of the spinal cord was achieved through field patching and computer optimization of the 3D fluence of each field. Sensitivity of the resultant plan to delivery and calculational errors was determined through both the assessment of the potential effects of range and patient setup errors, and by the application of Monte Carlo dose calculation methods. Ionization chamber profile measurements and 2D dosimetry using a scintillator/CCD camera arrangement were performed to verify the calculated fields in water. Modeling of a 10% overshoot of proton range showed that the maximum dose to the spinal cord remained unchanged, but setup error analysis showed that dose homogeneity in the target volume could be sensitive to offsets in the AP direction. No significant difference between the MC and analytic dose calculations was found and the measured dosimetry for all fields was accurate to 3% for all measured points. Over the course of the treatment, a setup accuracy of ±4 mm (2 s.d.) could be achieved, with a mean offset in the AP direction of 0.1 mm. Inhalation/exhalation CT scans indicated that organ motion in the region of the target volume was negligible. We conclude that 3D IMPT plans can be applied clinically and safely without modification to our existing delivery system. However, analysis of the calculated intensity matrices should be performed to assess the practicality, or otherwise, of the plan.Keywords
This publication has 15 references indexed in Scilit:
- Performance of a fluorescent screen and CCD camera as a two-dimensional dosimetry system for dynamic treatment techniquesMedical Physics, 2000
- Intensity modulation methods for proton radiotherapyPhysics in Medicine & Biology, 1999
- The precision of proton range calculations in proton radiotherapy treatment planning: experimental verification of the relation between CT-HU and proton stopping powerPhysics in Medicine & Biology, 1998
- The 200‐MeV proton therapy project at the Paul Scherrer Institute: Conceptual design and practical realizationMedical Physics, 1995
- X-ray field compensation with multileaf collimatorsInternational Journal of Radiation Oncology*Biology*Physics, 1994
- Application of fast simulated annealing to optimization of conformal radiation treatmentsMedical Physics, 1993
- Clinically relevant optimization of 3-D conformal treatmentsMedical Physics, 1992
- The generation of intensity-modulated fields for conformal radiotherapy by dynamic collimationPhysics in Medicine & Biology, 1992
- Calculation of the uncertainty in the dose delivered during radiation therapyMedical Physics, 1985
- Radiobiological studies of a high‐energy modulated proton beam utilizing cultured mammalian cellsCancer, 1975