Digital and conventional chest images: observer performance with Film Digital Radiography System.
- 1 January 1986
- journal article
- research article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 158 (1), 27-33
- https://doi.org/10.1148/radiology.158.1.3940392
Abstract
The Film Digital Radiography System (FilmDRS) is a device with a laser optical film digitizer, 2,000 .times. 2,000 .times. 12-bit memory, and a 1,000-line video display. To evaluate the adequacy of this device for general radiography of the chest, four readers independently analyzed both radiagraphs and the corresponding video display of the digitized chest images of 150 patients, consisting of 100 images of abnormalities and 50 normal images. The overall results indicate equal sensitivity for the two systems. The FilmDRS, with interactive windowing, proved superior in the detection of hilar and mediastinal disease. X-ray film was superior in allowing detection of hyperlucent states. There was equivalent sensitivity for other disease categories. Superior specificity was achieve with conventional radiographs.This publication has 7 references indexed in Scilit:
- Digital processing of film radiographsAmerican Journal of Roentgenology, 1985
- An on-line digital image management system.Radiology, 1984
- Computed radiography utilizing scanning laser stimulated luminescence.Radiology, 1983
- Scanned projection radiography of the chest versus standard film radiography: a comparison of 250 cases.Radiology, 1983
- Digital radiography of the chest: design features and considerations for a prototype unit.Radiology, 1983
- Diode array digital radiography: initial clinical experienceAmerican Journal of Roentgenology, 1982
- The effect of varying spatial resolution on the detectability of diffuse pulmonary nodules. Assessment with digitized conventional radiographs.Radiology, 1981