Skin dose and dose–area product values for interventional cardiology procedures
- 1 January 2001
- journal article
- research article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 74 (877), 48-55
- https://doi.org/10.1259/bjr.74.877.740048
Abstract
Coronary angiography and percutaneous transluminal coronary angioplasty procedures performed in four different facilities were monitored in the present study by measuring maximum skin dose, dose-area product and other operational parameters. Radiographic slow film, thermoluminescent dosemeters and transmission ion chambers were used to measure dose related quantities. Values of 107-711 mGy for maximum skin dose and 27.3-370.6 Gy cm2 for dose-area product were found, together with cumulative skin dose estimates of 110-3706 mGy. A discussion of the relationship of measured dose-area product and skin dose values is made using a field concentration factor defined as a way to interpret the findings. No general correlation was observed between dose-area product and maximum skin dose. Cumulative skin dose estimates throughout a procedure should be discarded as a realistic method for assessing deterministic risk in cardiology procedures. Slow film in addition to thermoluminescent dosemeters for measurement of maximum skin dose is a good alternative, especially for complex interventional procedures. For repeated procedures, combining film and dose-area product monitoring favours optimization of radiation protection for the patient.Keywords
This publication has 29 references indexed in Scilit:
- Severe Skin Reactions from Interventional Fluoroscopy: Case Report and Review of the LiteratureRadiology, 1999
- Entrance skin dose estimates derived from dose-area product measurements in interventional radiological procedures.The British Journal of Radiology, 1998
- Patient radiation doses during cardiac catheterization procedures.The British Journal of Radiology, 1998
- Radiation exposure to patients and operators during diagnostic catheterization and coronary angioplastyCatheterization and Cardiovascular Diagnosis, 1997
- Editorial comment: Radiation exposure in the cardiac catheterization laboratory: An issue or a non-issue?Catheterization and Cardiovascular Diagnosis, 1997
- Measurement of patient entrance surface dose rates for fluoroscopic X-ray unitsPhysics in Medicine & Biology, 1995
- Does the use of new intracoronary interventional devices prolong radiation exposure in the cardiac catheterization laboratory?Journal of the American College of Cardiology, 1994
- Potential Biological Effects Following High X-ray Dose Interventional ProceduresJournal of Vascular and Interventional Radiology, 1994
- Radiation risk to patients from percutaneous transluminal coronary angioplastyJournal of the American College of Cardiology, 1993
- Radiation exposure to patients undergoing percutaneous transluminal coronary angioplastyThe American Journal of Cardiology, 1987