Analysis of retrieved implants: Crystallinity changes in ultrahigh molecular weight polyethylene

Abstract
The extent of in vivo induced structural alterations were evaluated in 13 retrieved implants ranging from two weeks' to seven years' implantation. A statistically signifiant difference in crystallinity was seen between weight-bearing and non-weight-bearing region, regions of the implants. Further, in the weight-bearing region, crystallinity was higher in components implanted for two or more years than in those implanted for less than one-half year. These results were consistent with our prior in vitro study on the effects of sterilization and aging. Taken together, the data indicate that UHMW polyethylene is not a static material but is continually undergoing dynamic changes in vivo.

This publication has 8 references indexed in Scilit: