Errors in T2 estimation using multislice multiple‐echo imaging
- 1 January 1987
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 4 (1), 34-47
- https://doi.org/10.1002/mrm.1910040105
Abstract
Accurate T2 images calculated from multiple-echo sequences are difficult to obtain over a number of contiguous slices due to the presence of unwanted echoes that are generated at the slice edges. This problem is similar to problems encountered in single-slice imaging in the presence of rf pulse imperfections. In this paper, we use computer simulations to show that all unwanted echoes can be eliminated by defocusing them using additions to the slice-selection gradients. Accurate T2 images may then be calculated from the remaining echoes. We also present a method of experimentally displaying the slice shapes of unwanted echoes generated by a multislice imaging sequence. © 1987 Academic Press, Inc.This publication has 20 references indexed in Scilit:
- SHARP NMR Biomedical Spatial LocalizationMagnetic Resonance in Medicine, 1985
- Signal strength on a 0.15-T magnetic resonance imagerMedical Physics, 1985
- Exploiting the stimulated echo in nuclear magnetic resonance imaging. I. MethodJournal of Magnetic Resonance (1969), 1985
- Stimulated echo imagingJournal of Magnetic Resonance (1969), 1985
- Identification and characterization of tissues by T2‐selective whole‐body proton NMR imagingMagnetic Resonance in Medicine, 1985
- Variations in slice shape and absorption as artifacts in the determination of tissue parameters in NMR imagingMagnetic Resonance in Medicine, 1985
- Derivation of quantitative information in NMR imaging: a phantom studyPhysics in Medicine & Biology, 1984
- Potential problems with selective pulses in NMR imaging systemsMedical Physics, 1984
- Highly selective π2 and π pulse generationJournal of Magnetic Resonance (1969), 1984
- Errors in measurements of transverse relaxation ratesJournal of Magnetic Resonance (1969), 1973