Stability of Rapid Scanning Interferometers
- 1 September 1982
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 36 (5), 544-552
- https://doi.org/10.1366/0003702824639538
Abstract
An investigation is made to measure errors and instabilities associated with interferometric spectrometers. All data analysis is performed using the time or distance domain signal (interferogram) to eliminate any anomalies produced by transformation to spectral data. Three basic tests were performed which involved variations of the centerburst amplitude, the interferogram sampling offset and signal stability. These tests were performed on raw data and after all nonzero phase had been removed from the interferometric signal. All errors measured are extremely small, but are larger than ideal detector-limited noise. This paper does not attempt to identify all potential instabilities associated with interferometers, but presents methods by which instabilities can be measured and quantified on a relative basis.Keywords
This publication has 6 references indexed in Scilit:
- Quantitative Aspects of FT-IR Spectroscopy in Industrial ApplicationsPublished by Springer Nature ,1980
- The problem of nonlinear phase errors introduced by misalignment of a Michelson interferometerInfrared Physics, 1978
- Reconstruction of gas chromatograms from interferometric gas chromatography/infrared spectrometry dataAnalytical Chemistry, 1977
- Fourier Transform SpectroscopyPublished by Elsevier ,1972
- Correction of Asymmetric Interferograms Obtained in Fourier Spectroscopy*Journal of the Optical Society of America, 1966
- I. — les principes généraux des méthodes nouvelles en spectroscopie interférentielle - A propos de la théorie du spectromètre interférentiel multiplexJournal de Physique et le Radium, 1958