Evaluation of a Four-Channel Multiplexed Electrospray Triple Quadrupole Mass Spectrometer for the Simultaneous Validation of LC/MS/MS Methods in Four Different Preclinical Matrixes
- 15 March 2001
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 73 (8), 1740-1747
- https://doi.org/10.1021/ac0012694
Abstract
A four-channel multiplexed electrospray interface on a triple quadrupole mass spectrometer was evaluated for the simultaneous validation of LC/MS/MS methods for the quantitation of loratadine and its metabolite, descarboethoxyloratadine, in four different biological matrixes. The assays were performed in rat, rabbit, mouse, and dog plasma from 1 to 1000 ng/mL using 96-well solid-phase extraction for sample preparation. The limit of quantitation of 1 ng/mL corresponded to 5.56 pg of each analyte injected on-column. For the drug, quality control samples (n = 6 at four concentrations) had precision ranging from 0.967 to 16.0% and accuracy ranging from −8.44 to 10.5% across all four species. For the metabolite, the precision ranged from 0.684 to 11.0% and the accuracy was between 6.36 and −9.06%. Intersprayer cross talk for the multiplexed electrospray ion source was evaluated as a function of analyte concentration and was less than 0.08% at concentrations as high as 1000 ng/mL. These results demonstrate the feasibility of using parallel analysis to reduce the time required for method validation and to increase sample throughput in drug development studies.Keywords
This publication has 11 references indexed in Scilit:
- Bioanalytical High-Throughput Selected Reaction Monitoring-LC/MS Determination of Selected Estrogen Receptor Modulators in Human Plasma: 2000 Samples/DayAnalytical Chemistry, 2000
- High throughput liquid chromatography/mass spectrometric analyses using a novel multiplexed electrospray interfaceRapid Communications in Mass Spectrometry, 1999
- High-Throughput Bioanalytical LC/MS/MS Determination of Benzodiazepines in Human Urine: 1000 Samples per 12 HoursAnalytical Chemistry, 1999
- Direct plasma injection for high-performance liquid chromatographic–mass spectrometric quantitation of the anxiolytic agent CP-93 393Journal of Chromatography B: Biomedical Sciences and Applications, 1998
- Peer Reviewed: Sample Preparation for LC/MS/MS: Analyzing Biological and Environmental SamplesAnalytical Chemistry, 1998
- In vivo pharmacokinetic screening in cassette dosing experiments: the use of on-line prospekt® liquid chromatography/atmospheric pressure chemical ionization tandem mass spectrometry technology in drug discoveryRapid Communications in Mass Spectrometry, 1998
- High throughput liquid chromatography/mass spectrometry bioanalysis using 96-well disk solid phase extraction plate for the sample preparationRapid Communications in Mass Spectrometry, 1998
- A direct technique for the simultaneous determination of 10 drug candidates in plasma by liquid chromatography-atmospheric pressure chemical ionization mass spectrometry interfaced to a prospekt solid-phase extraction systemJournal of Pharmaceutical and Biomedical Analysis, 1997
- Rapid, Solid Phase Extraction Technique for the High-Throughput Assay of Darifenacin in Human PlasmaAnalytical Chemistry, 1996
- CYPRESS SYSTEMSAnalytical Chemistry, 1993