A SENSITIVE DILATOMETER FOR USE AT LOW TEMPERATURES
- 1 May 1964
- journal article
- Published by Canadian Science Publishing in Canadian Journal of Physics
- Vol. 42 (5), 835-847
- https://doi.org/10.1139/p64-079
Abstract
A new dilatometer for the measurement of thermal expansion at very low temperatures is described. It employs a doubly twisted strip which converts the linear expansion of a specimen into a rotation which can be measured with an optical lever. The present limit of detection of the apparatus is about 3 × 10−10 cm change in length of a specimen 8 cm long. So far, measurements have been made only on copper; these indicate that below 8 °K α is given by 3.10 × 10−11T3 + 2.78 × 10−10T/° K. This implies a limiting lattice Grueneisen constant, [Formula: see text], of 1.95, and an electronic γe of 1.21.Keywords