Ultrasonic nondiffracting transducer for medical imaging

Abstract
The nondiffracting J/sub 0/ Bessel beam is evaluated, and its application to medical imaging is suggested. Computer simulations and experimental results for a ten-ring annular Bessel shaded transducer are described. Both continuous-wave (CW) and pulse-wave (PW) excitations are shown and compared to conventional Gaussian beams. The nondiffracting beam has about 1.27-nm radius main lobe with a 20-cm depth of field compared to the Gaussian transducer of the same size with a 1.27-mm radius main lobe at a focus of 12 cm and 2*4-cm depth of field. The side lobes of the nondiffracting beam are the same as the J/sub 0/ Bessel function. The effects of heterogeneity due to tissue on the nondiffracting beam and on the focused Gaussian beam are also reported.<>