Heterophilic Antibodies Interfering With Radioimmunoassay
- 19 November 1982
- journal article
- research article
- Published by American Medical Association (AMA)
- Vol. 248 (19), 2489-2490
- https://doi.org/10.1001/jama.1982.03330190053031
Abstract
A young woman with amenorrhea had a consistently positive pregnancy test result (serum radioimmunoassay measurement of β-human chorionic gonadotropin hormone). No fetal or placental tissue was found after uterine curettage and exploratory laparotomy. The false-positive pregnancy test result was due to heterophilic antibovine and antigoat antibodies in the patient's serum. These antibodies interfered with radioimmunoassays using goat antibodies. This case shows that serum heterophilic antibodies can interfere with immunoassays and result in unnecessary diagnostic procedures and/or unnecessary treatment. (JAMA1982;248:2489-2490)Keywords
This publication has 3 references indexed in Scilit:
- Transient Neonatal Hyperthyrotropinemia: A Factitious Syndrome due to the Presence of Heterophilic Antibodies in the Plasma of Infants and Their MothersJournal of Clinical Endocrinology & Metabolism, 1981
- CIRCULATING ANTIBODIES TO OVINE AND BOVINE IMMUNOGLOBULIN IN HEALTHY SUBJECTS: A HAZARD FOR IMMUNOASSAYSThe Lancet, 1980
- [125I] protein A: A tracer for general use in immunoassayJournal of Immunological Methods, 1978