Reproducibility of relaxation times and spin density calculated from routine MR imaging sequences: clinical study of the CNS
- 1 June 1985
- journal article
- research article
- Published by American Roentgen Ray Society in American Journal of Roentgenology
- Vol. 144 (6), 1165-1170
- https://doi.org/10.2214/ajr.144.6.1165
Abstract
This study was undertaken to determine if routine clinical magnetic resonance imaging sequences using only two different repetition times (TRs) and with only two sequential spin echoes (SEs) can be used to calculate reproducible relaxation time and spin density values for normal central nervous system tissue using a 0.35 T production-model instrument. In 43 patients 650 regions of interest of 11 different anatomic sites were measured. T1 and T2 relaxation times and spin density were measured. For each anatomic location, the mean and standard deviation of these values were determined. In most solid regions of brain, the standard deviation of both T1 and T2 was 4%-8%. Relaxation times of cortical gray matter varied more, with a standard deviation of 10%, probably because of volume-averaging with adjacent cerebrospinal fluid (CSF). CSF and ocular vitreous humor were neither reproducibly nor accurately measured because of the short TR and TE settings of the imaging sequences relative to the long T1 and T2 rel...This publication has 4 references indexed in Scilit:
- Cerebral abnormalities: use of calculated T1 and T2 magnetic resonance images for diagnosis.Radiology, 1984
- Nuclear magnetic resonance: in vivo proton chemical shift imaging. Work in progress.Radiology, 1983
- Visualization of cerebral and vascular abnormalities by NMR imaging. The effects of imaging parameters on contrast.Radiology, 1982
- Explanation of Cerebral White-Gray Contrast in Computed TomographyJournal of Computer Assisted Tomography, 1980