The Development of the Electro-Magnetic Blood Flowmeter
- 1 January 1962
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IRE Transactions on Bio-Medical Electronics
- Vol. 9 (4), 228-235
- https://doi.org/10.1109/tbmel.1962.4323014
Abstract
Electromagnetic blood flowmeters were first constructed and used by Kolin in 1936 and, independently, by Wetterer in 1937. These early instruments consisted essentially of a large dc electromagnet, a pair of electrodes, and a string galvanometer; they made possible the recording of flow in exposed, large arteries of anesthetized animals under stringent experimental conditions. Modern electromagnetic flowmeters, using tiny, surgically implantable magnet-electrode assemblies and complex-electronic circuits, permit the recording of instantaneous and mean flows in large and small arteries in unanesthetized animals engaged in ordinary activities, as well as in anesthetized human patients whose vessels are exposed during surgical operations. Major steps in the development of the modern flowmeter include: 1) the early use of a sinusoidal ac-excited magnet to obviate the use of direct-coupled amplifiers and cumbersome, nonpolarizable electrodes; 2) the development of demodulation techniques for eliminating from the ac signal those induced voltages unrelated to flow: 3) the introduction of the squarewave and triangular-wave excited magnet to facilitate such signal-from-noise separation; 4) the miniaturization of magnetelectrode assemblies suitable for surgical implantation; and 5) improvements in circuitry resulting in greater electrical stability and freedom from noise.Keywords
This publication has 29 references indexed in Scilit:
- The DC Electromagnetic Flowmeter and Its Application to Blood Flow Measurement in Unopened VesselsIRE Transactions on Medical Electronics, 1959
- Flowmeter for Extracorporeal CirculationIRE Transactions on Medical Electronics, 1959
- Performance and Application of a Commercial Blood FlowmeterIRE Transactions on Medical Electronics, 1959
- Chopper-Operated Electromagnetic FlowmeterIRE Transactions on Medical Electronics, 1959
- Measurement of Cardiac Output in Unrestrained Dogs by an Implanted Electromagnetic MeterIRE Transactions on Medical Electronics, 1959
- Gated Sine-Wave Electromagnetic FlowmeterIRE Transactions on Medical Electronics, 1959
- A Magnetic Flowmeter for Recording Cardiac OutputProceedings of the IRE, 1959
- Electromagnetic Blood Flow MetersScience, 1959
- MINIATURIZATION OF THE ELECTROMAGNETIC BLOOD FLOW METER AND ITS USE FOR THE RECORDING OF CIRCULATORY RESPONSES OF CONSCIOUS ANIMALS TO SENSORY STIMULIProceedings of the National Academy of Sciences, 1959
- Electromagnetic Blood Flow Meter Yielding a Base Line Without Interruption of Flow.Experimental Biology and Medicine, 1958