Performance measurements of a depth-encoding PET detector module based on position-sensitive avalanche photodiode read-out
- 4 September 2004
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 49 (18), 4293-4304
- https://doi.org/10.1088/0031-9155/49/18/007
Abstract
We are developing a high-resolution, high-efficiency positron emission tomography (PET) detector module with depth of interaction (DOI) capability based on a lutetium oxyorthosilicate (LSO) scintillator array coupled at both ends to position-sensitive avalanche photodiodes (PSAPDs). In this paper we present the DOI resolution, energy resolution and timing resolution results for complete detector modules. The detector module consists of a 7 x 7 matrix of LSO scintillator crystals (1 x 1 x 20 mm3 in dimension) coupled to 8 x 8 mm2 PSAPDs at both ends. Flood histograms were acquired and used to generate crystal look-up tables. The DOI resolution was measured for individual crystals within the array by using the ratio of the signal amplitudes from the two PSAPDs on an event-by-event basis. A measure of the total scintillation light produced was obtained by summing the signal amplitudes from the two PSAPDs. This summed signal was used to measure the energy resolution. The DOI resolution was measured to be 3-4 mm FWHM irrespective of the position of the crystal within the array, or the interaction location along the length of the crystal. The total light signal and energy resolution was almost independent of the depth of interaction. The measured energy resolution averaged 14% FWHM. The coincidence timing resolution measured using a pair of identical detector modules was 4.5 ns FWHM. These results are consistent with the design goals and the performance required of a compact, high-resolution and high-efficiency PET detector module for small animal and breast imaging applications.Keywords
This publication has 22 references indexed in Scilit:
- Performance evaluation of the microPET R4 PET scanner for rodentsEuropean Journal of Nuclear Medicine and Molecular Imaging, 2003
- Detector development for microPET II: a 1 μl resolution PET scanner for small animal imagingPhysics in Medicine & Biology, 2001
- maxPET, a dedicated mammary and axillary region PET imaging system for breast cancerIEEE Transactions on Nuclear Science, 2001
- An LSO scintillator array for a PET detector module with depth of interaction measurementIEEE Transactions on Nuclear Science, 2001
- Development of a depth of interaction detector for γ-raysNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2001
- Conceptual design of a high-sensitivity small animal PET camera with 4/spl pi/ coverageIEEE Transactions on Nuclear Science, 1999
- Depth of interaction for PET using segmented crystalsIEEE Transactions on Nuclear Science, 1998
- Design of a depth of interaction (DOI) PET detector moduleIEEE Transactions on Nuclear Science, 1998
- Design of a high-resolution, high-sensitivity PET camera for human brains and small animalsIEEE Transactions on Nuclear Science, 1997
- MicroPET: a high resolution PET scanner for imaging small animalsIEEE Transactions on Nuclear Science, 1997