Visualization of flow patterns distal to aortic valve prostheses in humans using a fast approach for cine 3D velocity mapping
- 23 April 2001
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 13 (5), 690-698
- https://doi.org/10.1002/jmri.1097
Abstract
The fluid dynamic performance of mechanical heart valves differs from normal valves and thus is considered related to late clinical complications in patients. Since flow patterns evolving around heart valves are complex in space and time, flow visualization based on time‐resolved 3D velocity data might add important information regarding the performance of specific valve designs in vivo. However, previous cine 3D techniques for three‐directional phase‐contrast velocity mapping suffer from long scan duration and therefore might hamper assessment in patients. A hybrid 3D phase‐contrast sequence combining segmented k‐space acquisition with short EPI readout trains is presented with its validation in vitro. The technique was applied to study flow patterns downstream from a bileaflet aortic prosthesis in six patients. Navigator echoes were incorporated for respiratory motion compensation. Before flow visualization, spurious phase errors due to concomitant gradient fields and eddy currents were corrected. Flow visualization was based on particle paths and animated velocity vector plots. Dedicated algorithms for particle path integration were implemented to account for the considerable motion of the ascending aorta during the cardiac cycle. A distinct flow pattern reflecting the valve design was observed closest to the valve during early flow acceleration. Reverse flow occurred adjacent to high velocity jets and above the hinge housings. Later in systole, flow became confined to the central vessel area and reverse flow along the inner aortic curvature developed. Further downstream from the valve, flow patterns varied considerably among patients, indicating the impact of varying aortic anatomy in vivo. It is concluded that MR velocity mapping is a potential tool for studying 3D flow patterns evolving around heart valve prostheses in humans. J. Magn. Reson. Imaging 2001;13:690–698.Keywords
This publication has 16 references indexed in Scilit:
- Assessment of prosthetic aortic valve performance by magnetic resonance velocity imagingMagnetic Resonance Materials in Physics, Biology and Medicine, 2000
- 4D magnetic resonance velocity mapping of blood flow patterns in the aorta in young vs. elderly normal subjectsJournal of Magnetic Resonance Imaging, 1999
- Automatic vessel segmentation using active contours in cine phase contrast flow measurementsJournal of Magnetic Resonance Imaging, 1999
- Particle trace visualization of intracardiac flow using time-resolved 3D phase contrast MRIMagnetic Resonance in Medicine, 1999
- Multi-echo segmented k-space imaging: An optimized hybrid sequence for ultrafast cardiac imagingMagnetic Resonance in Medicine, 1999
- Visualizing blood flow patterns using streamlines, arrows, and particle pathsMagnetic Resonance in Medicine, 1998
- Concomitant gradient terms in phase contrast MR: Analysis and correctionMagnetic Resonance in Medicine, 1998
- Navigator-echo-based real-time respiratory gating and triggering for reduction of respiration effects in three-dimensional coronary MR angiography.Radiology, 1996
- Semiautomated method for noise reduction and background phase error correction in MR phase velocity dataJournal of Magnetic Resonance Imaging, 1993
- Velocity fields and turbulent stresses downstream of biological and mechanical aortic valve prostheses implanted in pigsCardiovascular Research, 1988