Wavelength Calibration of a Multichannel Spectrometer
- 1 November 1993
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 47 (11), 1808-1813
- https://doi.org/10.1366/0003702934065948
Abstract
A readily automated procedure for wavelength calibration of multichannel spectrometers is described. Once applied, line positions can be read from the multichannel display with good accuracy and precision for any spectrometer setting. The procedure uses Ne atomic lines as wavelength standards. A novel apodization procedure is used for accurate measurement of the pixel positions of the neon lines. Raman line positions can be determined with an average error of less than 0.2 cm−1.Keywords