Studies of a Glass-Ceramic Capacitance Thermometer in an Intense Magnetic Field at Low Temperatures
- 1 May 1971
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 42 (5), 571-573
- https://doi.org/10.1063/1.1685172
Abstract
The effect of an intense magnetic field on an 1100‐type capacitance thermometer (made from an SrTiO3 glass‐ceramic crystallized at 1100°C) was accurately measured at five temperatures between 1.5 and 4.2 K. At each manostatically maintained temperature, the capacitance changes were measured five times between B=0 and B=14 T, and it was found that these changes did not exceed the measurement uncertainty, ±0.3 pF≅±1 mK. Similar results were obtained on a second 1100‐type thermometer. Some suggestions are made for the use of this type of thermometer in magnetic field experiments.Keywords
This publication has 6 references indexed in Scilit:
- A Low Temperature Glass-Ceramic Capacitance ThermometerReview of Scientific Instruments, 1971
- Studies of a Glass-Ceramic Capacitance Thermometer between 0.025 and 2.4 KReview of Scientific Instruments, 1971
- Low Temperature Thermometry in High Magnetic Fields. II. Germanium and Platinum ResistorsReview of Scientific Instruments, 1971
- Cryogenic thermometry: a review of recent progressCryogenics, 1970
- Low Temperature Thermometry in High Magnetic Fields. I. Carbon ResistorsReview of Scientific Instruments, 1969
- Simple Mechanical Pressure Controller for Liquid-Helium CryostatsReview of Scientific Instruments, 1959