Segmental Analysis of Color Kinesis Images
- 15 April 1997
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation
- Vol. 95 (8), 2082-2097
- https://doi.org/10.1161/01.cir.95.8.2082
Abstract
Background We describe a method for objective assessment of left ventricular (LV) endocardial wall motion based on Color Kinesis, a new echocardiographic technique that color-encodes pixel transitions between blood and myocardial tissue. Methods and Results We developed a software that analyzes Color Kinesis images and provides quantitative indices of magnitude and timing of regional endocardial motion. Images obtained in 12 normal subjects were used to evaluate the variability in each index. Esmolol, dobutamine, and atropine were used to track variations in LV function in 14 subjects. Objective evaluation of wall motion was tested in 20 patients undergoing dobutamine stress testing. Regional fractional area change, displacement, and radial shortening were displayed as histograms and time curves. Global function was assessed by calculating magnitude and timing of peak ejection or filling rates and mean time of ejection or filling. Patterns of endocardial motion were consistent between normal subjects. Fractional area change and peak ejection rate decreased with esmolol and increased with dobutamine. Time to peak ejection and mean time of contraction were prolonged with esmolol and shortened with dobutamine. Using atropine, we proved that our findings with dobutamine were not secondary to its chronotropic effects. Dobutamine induced regional wall motion abnormalities in 10 patients in 38 segments diagnosed conventionally. Segmental analysis detected abnormalities in 36 of these 38 segments and in an additional 5 of 322 segments. Conclusions Analysis of Color Kinesis images allows fast, objective, and automated evaluation of regional wall motion sensitively enough to evaluate clinical dobutamine stress data. This method has significant potential in the diagnosis of myocardial ischemia.Keywords
This publication has 29 references indexed in Scilit:
- Stress Echocardiography. Part II. Dobutamine Stress Echocardiography: Techniques, Implementation, Clinical Applications, and CorrelationsMayo Clinic Proceedings, 1995
- Quantification of left ventricular function with an automated border detection system and comparison with radionuclide ventriculographyThe American Journal of Cardiology, 1994
- Comparison of left ventricular function by echocardiographic automated border detection and by radionuclide ejection fractionThe American Journal of Cardiology, 1993
- On-line estimation of changes in left ventricular stroke volume by transesophageal echpcardiographic automated border detection in patients undergoing coronary artery bypass graftingThe American Journal of Cardiology, 1993
- Wall motion asynchrony is a major determinant of impaired left ventricular filling in patients with healed myocardial infarctionThe American Journal of Cardiology, 1993
- Automated, on-line quantification of left ventricular dimensions and function by echocardiography with backscatter imaging and lateral gain compensationThe American Journal of Cardiology, 1992
- Influence of the orientation of myocardial fibers on echocardiographic imagesThe American Journal of Cardiology, 1987
- Heterogeneity of left ventricular segmental wall thickening and excursion in 2-dimensional echocardiograms of normal human subjectsThe American Journal of Cardiology, 1983
- Regional contraction patterns in the normal and ischemic left ventricle in manThe American Journal of Cardiology, 1973
- Dynamic geometry of the left ventricleThe American Journal of Cardiology, 1966