Instruments for Self-Monitoring of Blood Glucose: Comparisons of Testing Quality Achieved by Patients and a Technician
Open Access
- 1 July 2002
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 48 (7), 994-1003
- https://doi.org/10.1093/clinchem/48.7.994
Abstract
Background: Instruments for self-monitoring of blood glucose (SMBG) are increasingly used by patients with diabetes. The analytical quality of meters in routine use is poorly characterized. Methods: We compared SMBG performance achieved by patients and by a medical laboratory technician. Imprecision was calculated from duplicate measurements, and deviation as the difference between the first measurement and the mean of duplicate laboratory-method results (calibrated with NIST material). Analytical quality for five groups of SMBG instruments was compared with quality specifications for BG measurements. All participants completed a questionnaire assessing both SMBG training and use of the meters. Results: We recruited 159 SMBG users from a hospital outpatient clinic and 263 others from 65 randomly selected general practices (total of 422). Most (two thirds) used insulin. CVs for the five meter types were 7%, 11%, 18%, 18%, and 20% in the hands of patients and 2.5–5.9% for the technician. For three of five meter types, patients’ BG measurements had larger deviations from the laboratory results than did the technician’s results. The technician’s performance could not predict the patients’. No instrument when used by patients (but two operated by the technician) met published quality specifications. The analytical quality of patients’ results was not related to whether they had chosen the instruments on advice from healthcare personnel (one-third of patients), were only self-educated in SMBG (50%), or performed SMBG fewer than seven times/week (62%). Conclusions: The analytical quality of SMBG among patients was poorer than, and could not be predicted from, the performance of the meters in the hands of a technician. We suggest that new instruments be tested in the hands of patients who are trained on meter use in a routine way.This publication has 42 references indexed in Scilit:
- Self‐monitoring in Type 2 diabetes mellitus: a meta‐analysisDiabetic Medicine, 2000
- Intensive blood-glucose control with sulphonylureas or insulin compared with conventional treatment and risk of complications in patients with type 2 diabetes (UKPDS 33)The Lancet, 1998
- Effect of intensive blood-glucose control with metformin on complications in overweight patients with type 2 diabetes (UKPDS 34)The Lancet, 1998
- The Efficacy of Self-Monitoring of Blood Glucose in NIDDM subjects: A criteria-based literature reviewDiabetes Care, 1997
- Frequency and Impact of SMBG on Glycemic Control in Patients With NIDDM in an Urban Teaching Hospital ClinicThe Diabetes Educator, 1997
- Accuracy of home blood glucose meters during hypoglycemiaDiabetes Care, 1996
- What makes people with diabetes measure their own blood glucose?Practical Diabetes International, 1994
- Home glucose monitoring in patients aged over 40 years with diabetes mellitusPractical Diabetes International, 1994
- Home blood glucose monitoring: How often?Practical Diabetes International, 1993
- Home Monitors of Blood Glucose: Comparison of Precision and AccuracyDiabetes Care, 1987