Development of the Baylor/NASA Axial Flow Ventricular Assist Device: In Vitro Performance and Systematic Hemolysis Test Results
- 1 January 1994
- journal article
- Published by Wiley in Artificial Organs
- Vol. 18 (1), 32-43
- https://doi.org/10.1111/j.1525-1594.1994.tb03296.x
Abstract
Our newly developed axial flow pump consists of a flow tube, an internal rotating impeller, and a fixed flow stator (we call the stator) behind the impeller. This pump produces a flow of 3 to 8 L/min against 50 to 150 mm Hg pressure difference, respectively, in the range of 10,000 to 16,000 rpm. An axial flow pump that will be used as a ventricular assist device (VAD) has to have low hemolytic and good antithrombogenic characteristics. This paper will show how to decrease the hemolytic properties of this axial flow pump systematically using a test matrix. The test variables evaluated were impeller blade tip geometry, impeller flow tube clearance (radial clearance), impeller stator clearance (axial clearance), impeller blade number, stator blade number, and impeller length. All in vitro hemolysis tests were performed at 5.0 L/min against 100 mm Hg pressure difference using a total of 83 bags of fresh bovine blood. The results were as follows: the impeller blade tip geometry did not significantly effect hemolysis, a 0.005-inch and a 0.009-inch radial clearance were significantly (p < 0.01 or 0.001) less hemolytic than the other clearances, a 0.075-inch axial clearance was significantly (p < 0.05) more hemolytic than the other clearances, two- and six-bladed impellers were significantly (p < 0.01 and 0.02, respectively) less hemolytic than a four-bladed impeller, a five-bladed stator was significantly (p < 0.05 or 0.01) less hemolytic than the other stators, and the impeller length did not make a significant difference. Currently, the best index of hemolysis is 0.031 +/- 0.018 g/100 L, and using parameters from these results, implantable devices are being fabricated.Keywords
This publication has 12 references indexed in Scilit:
- In Vitro Performance of the Baylor/NASA Axial Flow PumpArtificial Organs, 1993
- Left ventricular assist with an axial flow pump (Hemopump): Clinical applicationThe Annals of Thoracic Surgery, 1993
- Development of an Atraumatic Small Centrifugal Pump for Second‐Generation Cardiopulmonary BypassArtificial Organs, 1992
- A Compact Centrifugal Pump for Cardiopulmonary BypassArtificial Organs, 1992
- Development of an Axial Flow Blood Pump LVASAsaio Journal, 1992
- The Valvo‐Pump, An Axial Blood Pump Implanted at the Heart Valve Position: Concept and Initial ResultsArtificial Organs, 1992
- Treatment of cardiogenic shock with the hemopump left ventricular assist deviceThe Annals of Thoracic Surgery, 1991
- The Spindle Pump—A Nonpulsatile Blood Pump for Assisted CirculationJournal of Biomaterials Applications, 1990
- The Need for a Nonpulsatile Pumping SystemArtificial Organs, 1988
- Contribution to the in vitro testing of pumps for extracorporeal circulationThe Journal of Thoracic and Cardiovascular Surgery, 1967