Prospectively gated cardiac computed tomography

Abstract
A fourth-generation scanner has been modified to perform prospectively gated cardiac computed tomography (CT). A computer program monitors the electrocardiogram (ECG) and predicts when to initiate the next scan in a gated series in order to acquire all projection data for a desired phase of the heart cycle. The system has been tested with dogs and has produced cross-sectional images of all phases of the cardiac cycle. Eight to ten scans per series were sufficient to obtain reproducible images of each transverse section in the end-diastolic and end-systolic phases. The radiation dose to the skin was approximately 1.4 cGy per scan. The prospectively gated system is more than twice as efficient as a retrospectively gated system in obtaining complete angular projection data for a 10% heart cycle window. A temporal smoothing technique to suppress reconstruction artifacts due to sorting inconsistent projection data was developed and evaluated. Image noise was reduced by averaging together any overlapping projection data. Prospectively gated cardiac CT has also been used to demonstrate that the error in attenuation measured with a single nongated CT scan through the heart can be as large as 50-60 CT numbers outside the heart in the lung field.