Validation of a Monte Carlo simulation of the Philips Allegro/GEMINI PET systems using GATE
- 1 February 2006
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 51 (4), 943-962
- https://doi.org/10.1088/0031-9155/51/4/013
Abstract
A newly developed simulation toolkit, GATE (Geant4 Application for Tomographic Emission), was used to develop a Monte Carlo simulation of a fully three-dimensional (3D) clinical PET scanner. The Philips Allegro/GEMINI PET systems were simulated in order to (a) allow a detailed study of the parameters affecting the system's performance under various imaging conditions, (b) study the optimization and quantitative accuracy of emission acquisition protocols for dynamic and static imaging, and (c) further validate the potential of GATE for the simulation of clinical PET systems. A model of the detection system and its geometry was developed. The accuracy of the developed detection model was tested through the comparison of simulated and measured results obtained with the Allegro/GEMINI systems for a number of NEMA NU2-2001 performance protocols including spatial resolution, sensitivity and scatter fraction. In addition, an approximate model of the system's dead time at the level of detected single events and coincidences was developed in an attempt to simulate the count rate related performance characteristics of the scanner. The developed dead-time model was assessed under different imaging conditions using the count rate loss and noise equivalent count rates performance protocols of standard and modified NEMA NU2-2001 (whole body imaging conditions) and NEMA NU2-1994 (brain imaging conditions) comparing simulated with experimental measurements obtained with the Allegro/GEMINI PET systems. Finally, a reconstructed image quality protocol was used to assess the overall performance of the developed model. An agreement of <3% was obtained in scatter fraction, with a difference between 4% and 10% in the true and random coincidence count rates respectively, throughout a range of activity concentrations and under various imaging conditions, resulting in <8% differences between simulated and measured noise equivalent count rates performance. Finally, the image quality validation study revealed a good agreement in signal-to-noise ratio and contrast recovery coefficients for a number of different volume spheres and two different (clinical level based) tumour-to-background ratios. In conclusion, these results support the accurate modelling of the Philips Allegro/GEMINI PET systems using GATE in combination with a dead-time model for the signal flow description, which leads to an agreement of <10% in coincidence count rates under different imaging conditions and clinically relevant activity concentration levels.Keywords
This publication has 7 references indexed in Scilit:
- Evaluation of attenuation correction methodology in the allegro PET systemIEEE Transactions on Nuclear Science, 2004
- GATE: a simulation toolkit for PET and SPECTPhysics in Medicine & Biology, 2004
- Geant4—a simulation toolkitNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2003
- A comparison of segmentation and emission subtraction for singles transmission in PETIEEE Transactions on Nuclear Science, 1998
- ROOT — An object oriented data analysis frameworkNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1997
- Measuring PET scanner sensitivity: relating countrates to image signal-to-noise ratios using noise equivalents countsIEEE Transactions on Nuclear Science, 1990
- Evemt Localization in a Continuous Scintillation Detector Using Digital ProcessingIEEE Transactions on Nuclear Science, 1986