Serum ionised calcium concentration: measurement versus calculation.
- 29 April 1978
- Vol. 1 (6120), 1103-1105
- https://doi.org/10.1136/bmj.1.6120.1103
Abstract
Four hundred and eighteen measurements of serum ionised calcium, total calcium, and protein concentrations were made from 47 normal volunteers, 104 patients with chronic renal failure (33 being treated conservatively and 71 with regular haemodialysis), and 83 renal transplant recipients. The serum ionised calcium concentration was measured with an Orion SS-20 meter and calculated from the total serum calcium and protein concentrations by using three formulae and a nomogram. In the normal subjects and patients undergoing regular haemodialysis, whose serum calcium concentrations were in or near the normal range, three of the calculations gave results similar to those obtained by direct measurement. In patients with conservatively treated chronic renal failure and those who had received renal transplants, however, there was poor aggrement between the methods. When patients with hypercalcaemia and hypocalcaemia from all the groups were considered separately there was again poor agreement between calculated and measured concentrations of serum ionised calcium. Of the patients whose measured concentrations of serum ionised calcium were high, 69-76% were classified as normal by the four indirect methods. We conclude that calculation of the serum ionised calcium concentrations is not an adequate substitute for direct measurement.Keywords
This publication has 6 references indexed in Scilit:
- Measurement of serum and plasma ionic calcium with the "Space-Stat 20 Ionized Calcium Analyzer".Clinical Chemistry, 1977
- EVALUATION OF A NEW AUTOMATIC CALCIUM-ION ANALYZER1977
- Letter: Adjustment of plasma calcium measurements.BMJ, 1974
- Variation in Plasma Calcium with Induced Changes in Plasma Specific Gravity, Total Protein, and AlbuminBMJ, 1973
- Determination of Diffusible Serum CalciumAmerican Journal of Clinical Pathology, 1954
- Construction of a Cartesian Nomogram for the Law of Mass ActionScience, 1935