Sources of Error in Glucose Determinations in Neonatal Blood by Glucose Oxidase Methods, Including Dextrostix

Abstract
An investigation was carried out to examine why a glucose oxidase-peroxidase-orthodianisidine method for plasma glucose, without protein precipitation, gave low results for neonatal blood. The magnitude of the difference between the results with and without protein precipitation was examined in a clinical neonatal series, and in sera to which bilirubin, hemolysate, pure hemoglobin, and uric acid had been added. Systematic linear inhibition was demonstrated with bilirubin, and the results suggested that high concentrations of hemolysate and uric acid could also interfere. Use of alkaline protein precipitants eliminated the inhibition. Dextrostix test results for neonatal blood are compared with results of conventional glucose analyses and possible sources of discrepancy examined.