Dual Chamber Arrhythmia Detection in the Implantable Cardioverter Defibrillator
- 1 October 2000
- journal article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 11 (10), 1105-1115
- https://doi.org/10.1111/j.1540-8167.2000.tb01756.x
Abstract
Dual Chamber Arrhythmia Detection hy ICD. Introduction. Dual chamber implantable cardioverter detibrillator (ICD) technology extended ICD therapy to more than termination of hemodynamicaly unstable ventricular tachyarrhythmias. It created the basis for dual chamber arrhythmoa management. In which dependable detection is important for treatment and prevention of both ventricular and atrial arrhythmias. Methods and Results. Dual chamber detection algorithms were investigated in two Medironic dual chamber ICDs; the 7250 Jewel AF (33 patients) and the 7217 Gem DR (31 patients). Both ICDS use the same PR Logic algorithm to interpret tachycardia as ventricular tachycardia (VT), supraventricular tachycardia (SVT), or dual (VT+SVT), The accuracy of dual chamber detection was studied in 310 of 1,367 spontaneously occurring tachcardias in which rate criterion only was not sufficient for arrhythmia diagnosis. In 78 epidoses there was a double tachycardia, in 223 episodes. SVT was detected in the VT or veutricular fibrillation zone, and in 9 episodes arrhythmia was detected outside the boundaries of the PR Logic functioning. In 100% of double techycardias the VT was correctly diagonsed and reveived priority treatment. SVT was seen in 59 (19%) episodes diagonsed as VT. The causes of inappropriate detection were (1) algorithm failure (inability to fulfill the PRJ Cardiovasc Electrophysiol. Vot. 11. pp. 1105‐1115. October 2000)Keywords
This publication has 17 references indexed in Scilit:
- Enhanced Detection Criteria in Implantable DefibrillatorsJournal of Cardiovascular Electrophysiology, 1998
- Ventriculoatrial Conduction Capability and Prevalence of 1:1 Retrograde Conduction During Inducible Sustained Monomorphic Ventricular Tacbycardia in 305 Implantable Cardioverter Defibrillator RecipientsPacing and Clinical Electrophysiology, 1997
- Analysis of Atrial Sensed Far‐Field Ventricnlar Signals: A ReassessmentPacing and Clinical Electrophysiology, 1997
- Failure of Automatic Mode Switching: Recognition and ManagementPacing and Clinical Electrophysiology, 1997
- Programmable VT Detection Enhancements in Implantable Cardioverter Defibrillator TherapyPacing and Clinical Electrophysiology, 1995
- Underdetection of ventricular tachycardia by algorithms to enhance specificity in a tiered-therapy cardioverter-defibrillatorJournal of the American College of Cardiology, 1994
- Holter Documented Sudden Death in a Patient with an Implanted DefibrillatorPacing and Clinical Electrophysiology, 1992
- Automatic Recognition of Ventricular Arrhythmias Using Temporal Electrogram AnalysisPacing and Clinical Electrophysiology, 1991
- The Use of Exercise Testing in Children to Evaluate Abnormalities of Pacemaker Function not Apparent at RestPacing and Clinical Electrophysiology, 1985