Calibration of Infra-red CO2 Gas Analyzers
Open Access
- 1 August 1969
- journal article
- Published by Oxford University Press (OUP) in Plant Physiology
- Vol. 44 (8), 1122-1126
- https://doi.org/10.1104/pp.44.8.1122
Abstract
Precision gas mixing pumps produce CO2 gas mixtures for the calibration of infra-red CO2 gas analyzers equivalent in accuracy to the standard CO2 gas mixtures (± 1%) supplied by the National Bureau of Standards, Washington, D. C. The calibration of infra-red gas analyzers by the pressure difference method and by concentration differences did not agree. A factor of 0.709 was necessary to correct the pressure difference method for the effects of pressure broadening with our analyzer.Keywords
This publication has 1 reference indexed in Scilit: