Reproducibility of cardiac output measurement by cross sectional and Doppler echocardiography.
- 1 June 1988
- Vol. 59 (6), 680-684
- https://doi.org/10.1136/hrt.59.6.680
Abstract
The variability of Doppler echocardiographic estimation of cardiac output at the aortic orifice was investigated in eight healthy subjects. Cross sectional echocardiograms of the aortic orifice and aortic Doppler velocities were recorded and measured by four echocardiographers. Between subject variability was significantly larger than within subject variability for all variables. Variability owing to different echocardiographers and different measurement times was small compared with total variability. Coefficients of variation for aortic annular diameter, aortic velocity integral, and heart rate were 4.1%, 6.4%, and 5.0% respectively. The coefficient of variation for cardiac output was 8.8% and the 95% confidence interval for measurement of cardiac output by the Doppler method was 4.45 to 6.35 l/min. One echocardiographer reanalysed all the recordings and the results showed that recording the echocardiograms introduces a significantly larger source of error than measuring them. Thus serial measurements of cardiac output by the Doppler method can be performed with acceptable reproducibility; this indicates that the technique can be accurately applied in clinical practice.Keywords
This publication has 18 references indexed in Scilit:
- Reproducibility of cardiac stroke volume estimated by doppler echocardiographyThe American Journal of Cardiology, 1987
- Limits of reproducibility of blood flow measurements by doppler echocardiographyThe American Journal of Cardiology, 1987
- Accuracy of determination of changes in cardiac output by transcutaneous continuous-wave doppler computerThe American Journal of Cardiology, 1984
- Reproducibility of doppler aortic blood flow measurements: Studies on intraobserver, interobserver and day-to-day variability in normal subjectsThe American Journal of Cardiology, 1984
- Pulsed Doppler echocardiographic determination of stroke volume and cardiac output: clinical validation of two new methods using the apical window.Circulation, 1984
- Noninvasive Measurement of Cardiac Output by Continuous-Wave Doppler Echocardiography: Initial Experience and Review of the LiteratureMayo Clinic Proceedings, 1984
- Determination of cardiac output by Doppler echocardiography.Heart, 1984
- Doppler echocardiographic measurement of cardiac output: Comparison with a non-golden standardThe American Journal of Cardiology, 1984
- Determination of cardiac output by transcutaneous continuous-wave ultrasonic Doppler computerThe American Journal of Cardiology, 1984
- Noninvasive Doppler determination of cardiac output in man. Clinical validation.Circulation, 1983