Quantifying the MB isoenzyme of creatine kinase with the Abbott "IMx" immunoassay analyzer
- 1 February 1990
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 36 (2), 375-378
- https://doi.org/10.1093/clinchem/36.2.375
Abstract
In this two-step automated assay of the MB isoenzyme of creatine kinase (CK-MB), developed for the Abbott "IMx" immunoassay analyzer, monoclonal anti-CK-MB antibody immobilized onto latex microparticles and polyclonal anti-CK-MM antibody coupled to alkaline phosphatase are used. Within-run CVs ranged from 3.9% to 9.0%, between-run CVs from 0.0% to 5.6%, and the sensitivity was 0.2 microgram/L. Twenty-four results can be obtained in about 37 min. Analytical recovery of CK-MB added to human serum or plasma ranged from 89% to 109%. Icteric, lipemic, or hemolyzed samples did not interfere with CK-MB recovery. Cross-reactivity with CK-MM and CK-BB was 0.012% and 0.001%, respectively. The normal reference interval was 0-5 micrograms/L. IMx CK-MB results correlated well with CK-MB enzyme activity as determined by electrophoresis (n = 203; r = 0.97; slope = 0.90; y-intercept = -4.29) and with commercial immunoassays. We think that this assay will be useful for confirmation of acute myocardial infarction, both in critical-care units and in the clinical laboratory.This publication has 1 reference indexed in Scilit: