Catheter Embolization from Implanted Venous Access Devices: Case Reports
- 1 April 1989
- journal article
- case report
- Published by SAGE Publications in Angiology
- Vol. 40 (4), 319-323
- https://doi.org/10.1177/000331978904000413
Abstract
Two cases of catheter embolization from implanted venous access devices are reported and the available literature is reviewed. The catheter from an im planted venous access device migrated into the right heart after slippage of the O-ring, which attaches the catheter to the infusion port. The distal 6 cm of an infusion port catheter embolized to the right heart after spontaneous fracture of the catheter at the point where it passed between the clavicle and first rib. Both catheters were removed percutaneously without complication. Risk factors for embolization were apparent on x-ray films with evidence of O-ring slippage in 1 case an obvious kinking of the catheter in the other. Symptoms of embolization included chest discomfort, right upper quadrant pain, and nausea. In 1 case, an extra heart sound, initially thought to be an S3, disappeared when the catheter was removed.Keywords
This publication has 6 references indexed in Scilit:
- Spontaneous intravenous catheter fracture and embolization from an implanted venous access port and analysis by scanning electron microscopyCancer, 1987
- Hickman Catheter SeparationJournal of Parenteral and Enteral Nutrition, 1985
- Complications and management of implanted venous access catheters.Journal of Clinical Oncology, 1985
- Major complications of percutaneous subclavian vein cathetersThe American Journal of Surgery, 1979
- Management of embolised central venous catheters.BMJ, 1978
- Evaluation of current techniques for nonsurgical removal of intravascular iatrogenic foreign bodiesAmerican Journal of Roentgenology, 1978