Corrections of Humidity Measurement Errors from the Vaisala RS80 Radiosonde—Application to TOGA COARE Data

Abstract
A series of laboratory tests have been conducted on several different batches of Vaisala RS80 radiosondes to understand and develop methods to correct six humidity measurement errors, including chemical contamination, temperature dependence, basic calibration model, ground check, sensor aging, and sensor arm heating. The contamination and temperature-dependence (TD) errors dominate total errors. The chemical contamination error produces a dry bias, and is due to the occupation of binding sites in the sensor polymer by nonwater molecules emitted from the sonde packaging material. The magnitude of the dry bias depends on sensor polymer type (RS80-A and RS80-H), age of the sonde, relative humidity (RH), and temperature, and it exists throughout the troposphere. The contamination error generally increases with age and RH, and is larger for the RS80-H than the RS80-A. It is ∼2% and ∼10% at saturation for 1-yr-old RS80-A and RS80-H sondes, respectively. The TD error for the RS80-A results from an appro... Abstract A series of laboratory tests have been conducted on several different batches of Vaisala RS80 radiosondes to understand and develop methods to correct six humidity measurement errors, including chemical contamination, temperature dependence, basic calibration model, ground check, sensor aging, and sensor arm heating. The contamination and temperature-dependence (TD) errors dominate total errors. The chemical contamination error produces a dry bias, and is due to the occupation of binding sites in the sensor polymer by nonwater molecules emitted from the sonde packaging material. The magnitude of the dry bias depends on sensor polymer type (RS80-A and RS80-H), age of the sonde, relative humidity (RH), and temperature, and it exists throughout the troposphere. The contamination error generally increases with age and RH, and is larger for the RS80-H than the RS80-A. It is ∼2% and ∼10% at saturation for 1-yr-old RS80-A and RS80-H sondes, respectively. The TD error for the RS80-A results from an appro...