International Multicenter Evaluation of Latex Agglutination Tests for Identification of Staphylococcus aureus

Abstract
A newly marketed rapid agglutination kit for the identification of Staphylococcus aureus , Slidex Staph Plus (bioMérieux), was compared to Staphaurex Plus (Murex Diagnostics) and Pastorex Staph-Plus (Sanofi Diagnostics Pasteur). The study took place in three clinical microbiology laboratories in three different European countries. A total of 892 staphylococcal isolates, including 278 methicillin-sensitive S. aureus (MSSA) isolates, 171 methicillin-resistant S. aureus (MRSA) isolates, and 443 coagulase-negative staphylococcal isolates, were analyzed. The sensitivities (MSSA/MRSA) and specificities, respectively, were 98.2% (98.9%/97.1%) and 98.9% for Slidex Staph Plus, 98.2% (98.2%/98.2%) and 96.2% for Staphaurex Plus, and 98.7% (98.6%/98.8%) and 95.7% for Pastorex Staph Plus. The specificity of the Slidex Staph Plus kit was statistically significantly higher than the specificities of Staphaurex Plus and Pastorex Staph-Plus. The Slidex Staph Plus is a very reliable test for the identification of S. aureus .