Abstract
Cadmium, cobalt, and nickel at ng/g to μg/g levels in plant and animal tissue reference materials and at μg/L levels in blood and urine were determined by flame atomic absorption spectrometry. The analyte elements were preconcentrated and separated from the bulk of the matrix by on-line co-precipitation with the hexamethylene ammonium hexamethylene dithiocarbamate iron(II) chelate complex in a flow injection system. The precipitate was collected in a knotted reactor made from 150-cm-long, 0.5-mm-i.d. Microline tubing without using a filter. The precipitate was dissolved in methyl isobutyl ketone and introduced directly into the nebulizer-burner system of an atomic absorption spectrometer. Ascorbic acid in an HC1/KC1 buffer was added on-line in order to reduce iron(III) to iron(II) because of its much better efficiency as a collector for trace elements. Reagent concentrations were optimized so that at least 200 mg/L of iron and 15 mg/L of copper could be tolerated in the sample solution without causing significant interferences. The portion of the analyte retained in the collector was about 70% for cadmium and 50% for cobalt and nickel. Enrichment factors of 24, 19, and 20 were obtained for cadmium, cobalt, and nickel, respectively, with the use of a 40-s co-precipitation time, resulting in enhancement factors, including the effect of the organic solvent, of 52, 43, and 52, respectively. The detection limits (3σ) for cadmium, cobalt and nickel were 0.15, 1.3, and 1.5 μg/L, respectively, and the precision was 1.5% RSD for 10 μg/L Cd, 2.7% RSD for 50 μg/L Co, and 1.8% RSD for 50 μg/L Ni. The analytical results obtained for a number of standard reference materials and control samples were in good agreement with the certified or recommended values.Keywords
This publication has 8 references indexed in Scilit:
- Determination of gold in ore by flame atomic absorption spectrometry with flow-injection on-line sorbent extraction preconcentrationAnalytica Chimica Acta, 1991
- Determination of cadmium, lead and copper in water samples by flame atomic-absorption spectrometry with preconcentration by flow-injection on-line sorbent extractionTalanta, 1991
- Flame atomic absorption spectrometric determination of lead in biological samples using a flow injection system with on-line preconcentration by coprecipitation without filtrationJournal of Analytical Atomic Spectrometry, 1991
- High efficiency low sample consumption on-line ion-exchange pre-concentration system for flow injection flame atomic absorption spectrometryJournal of Analytical Atomic Spectrometry, 1989
- Determination of lead in whole blood by graphite furnace atomic absorption spectrometry with matrix modificationJournal of Analytical Atomic Spectrometry, 1989
- Zur Multi-Elementanreicherung aus Reineisen und reinen EisenverbindungenAnalytical and Bioanalytical Chemistry, 1988
- Zur Multi-Elementanreicherung aus eisenhaltigen Böden und SedimentenAnalytical and Bioanalytical Chemistry, 1987
- Determination of lead in blood using electrothermal atomisation atomic absorption spectrometry with a l'vov platform and matrix modifierThe Analyst, 1987