An undetectable source of technical error that could lead to false negative results in enzyme linked immunosorbent assay of antibodies to HIV‐1
- 1 January 1989
- journal article
- research article
- Published by Wiley in Transfusion
- Vol. 29 (1), 75-77
- https://doi.org/10.1046/j.1537-2995.1989.29189101170.x
Abstract
Since the institution of routine testing for antibodies to Human Immunodeficiency Virus (HIV) using the enzyme-linked immunosorbent assay (ELISA) the specificity and sensitivity of this assay system has received significant scrutiny. During previous use of this methodology, we have quantified rates of false biological positive results using commercial kit assays in a normal donor population. In this study, we have identified a potential source for false negative results. Using multiple lots of two different commercial ELISA kits, the absorbance readings at the test end point could not differentiate between normal non-reactive donor samples and blanks containing no sample. Thise results occur using normal donor samples even though the assays could distinguish between blank wells and the manufacturers'' "normal controls", provided with the assay. Our findings suggest that a technical pipetting error is presently undetectable, either visually or by statistical methods, and could permit an untested, potentially HIV-1 positive, unit to be released into the transfusable blood supply. A possible solution is suggested.This publication has 9 references indexed in Scilit:
- Lack of evidence of prolonged human immunodeficiency virus infection before antibody seroconversionBlood, 1988
- Human immunodeficiency virus test evaluation, performance, and use. Proposals to make good tests betterJAMA, 1988
- Transfusion-Transmitted AIDS ReassessedNew England Journal of Medicine, 1988
- Transmission of Human Immunodeficiency Virus (HIV) by Blood Transfusions Screened as Negative for HIV AntibodyNew England Journal of Medicine, 1988
- Estimating the risks of transfusion‐associated acquired immune deficiency syndrome and human immunodeficiency virus infectionTransfusion, 1987
- LONG LATENCY PRECEDES OVERT SEROCONVERSION IN SEXUALLY TRANSMITTED HUMAN-IMMUNODEFICIENCY-VIRUS INFECTIONThe Lancet, 1987
- Transfusion-Associated Hepatitis and AIDSNew England Journal of Medicine, 1987
- A second discriminator for biological false positive results in enzyme‐ linked immunosorbent assays for antibodies to human immunodeficiency virus (HTLV‐III/LAV)Transfusion, 1987
- The Impact of Routine HTLV-III Antibody Testing of Blood and Plasma Donors on Public HealthPublished by American Medical Association (AMA) ,1986