Survey of the optical quality of intraocular lens implants
- 1 March 1991
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 17 (2), 168-174
- https://doi.org/10.1016/s0886-3350(13)80247-7
Abstract
Intraocular lenses (IOLs) from 15 U.S. firms were tested for conformance to the requirements of the American National Standards Institute Z80.7 standard for IOLs. A total of 162 IOLs were tested for resolution, astigmatism, and accuracy of labeled power. Average resolving power was 78% of the diffraction limit, much better than the ANSI minimum requirement which is typically equivalent to 30% of the diffraction limit. This suggests that the ANSI Z80.7 requirement could be significantly tightened with little effect on current production practices. Only one IOL exhibited astigmatism in excess of 0.25 diopter. Differences between measured and labeled power in excess of 0.50 diopter were found in 22 lenses, indicating that accuracy of refractive power may be the most commonly missed optical requirement of the standard.Keywords
This publication has 10 references indexed in Scilit:
- Resolution testing of intraocular lensesJournal of Cataract & Refractive Surgery, 1991
- Silicone intraocular lens resolution in air and in waterJournal of Cataract & Refractive Surgery, 1988
- Cataract, intraocular lens, and refractive surgery in 1987 with a forecast to 1995Journal of Cataract & Refractive Surgery, 1988
- Intraocular lens resolution in air and waterJournal of Cataract & Refractive Surgery, 1987
- Visual aberrations caused by optic components of posterior chamber intraocular lensesJournal of Cataract & Refractive Surgery, 1987
- Unwanted optical effects caused by intraocular lens positioning holesJournal of Cataract & Refractive Surgery, 1987
- Optical Aberrations From a Well-Centered Intraocular Lens ImplantAmerican Journal of Ophthalmology, 1986
- Intraocular lens optical quality: update 1979American Intra-Ocular Implant Society Journal, 1980
- The Optical Quality of Currently Manufactured Intraocular LensesAmerican Journal of Ophthalmology, 1979
- The resolving power of intraocular lens implantsAmerican Intra-Ocular Implant Society Journal, 1978