Calibration of rectangular atomic force microscope cantilevers
- 1 October 1999
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 70 (10), 3967-3969
- https://doi.org/10.1063/1.1150021
Abstract
The spring constant of an atomic force microscope cantilever is often needed for quantitative measurements. The calibration method of Sader et al. [Rev. Sci. Instrum. 70, 3967 (1999)]10.1063/1.1150021 for a rectangular cantilever requires measurement of the resonant frequency and quality factor in fluid (typically air), and knowledge of its plan view dimensions. This intrinsically uses the hydrodynamic function for a cantilever of rectangular plan view geometry. Here, we present hydrodynamic functions for a series of irregular and non-rectangular atomic force microscope cantilevers that are commonly used in practice. Cantilever geometries of arrow shape, small aspect ratio rectangular, quasi-rectangular, irregular rectangular, non-ideal trapezoidal cross sections, and V-shape are all studied. This enables the spring constants of all these cantilevers to be accurately and routinely determined through measurement of their resonant frequency and quality factor in fluid (such as air). An approximate formulation of the hydrodynamic function for microcantilevers of arbitrary geometry is also proposed. Implementation of the method and its performance in the presence of uncertainties and non-idealities is discussed, together with conversion factors for the static and dynamic spring constants of these cantilevers. These results are expected to be of particular value to the design and application of micro- and nanomechanical systems in generalKeywords
This publication has 11 references indexed in Scilit:
- Frequency response of cantilever beams immersed in viscous fluids with applications to the atomic force microscopeJournal of Applied Physics, 1998
- Characterization of application-specific probes for SPMsPublished by SPIE-Intl Soc Optical Eng ,1997
- Short cantilevers for atomic force microscopyReview of Scientific Instruments, 1996
- Parallel beam approximation for V-shaped atomic force microscope cantileversReview of Scientific Instruments, 1995
- Method for the calibration of atomic force microscope cantileversReview of Scientific Instruments, 1995
- Experimental Determination of Spring Constants in Atomic Force MicroscopyLangmuir, 1994
- Theoretical analysis of the static deflection of plates for atomic force microscope applicationsJournal of Applied Physics, 1993
- Calibration of atomic-force microscope tipsReview of Scientific Instruments, 1993
- Direct measurement of interaction forces between colloidal particles using the scanning force microscopeLangmuir, 1993
- A nondestructive method for determining the spring constant of cantilevers for scanning force microscopyReview of Scientific Instruments, 1993