Electron-Beam Arc Therapy Using a High Energy Betatron
- 1 November 1979
- journal article
- research article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 133 (2), 483-489
- https://doi.org/10.1148/133.2.483
Abstract
Dose distributions in electron-beam arc [cancer] therapy were investigated using the electron beam from a Brown Boveri 45 MeV betatron. This variable isocenter machine is equipped with an angular speed control device that synchronizes the radiation dose rate with the rotation speed about the isocenter. An Alderson Rando phantom was used with Kodak RP/V film to measure isodensity distributions that were then converted to isodose curves. The effects of varying field size, isocenter depth, and air gap between collimator and phantom were examined, along with changes in the dose distributions produced by wax bolus in part of the arc.This publication has 1 reference indexed in Scilit:
- Physical Aspects of Electron-Beam Arc TherapyRadiology, 1977