Echo‐planar imaging of diffusion and perfusion
- 1 June 1991
- journal article
- review article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 19 (2), 247-253
- https://doi.org/10.1002/mrm.1910190210
Abstract
Use of the Stejskal‐Tanner sequence for performing diffusion images in the human brain tends to be complicated by the presence of artifacts caused by voluntary or involuntary, sometimes pulsatile, motion. We describe the implementation of the technique of echo‐planar diffusion imaging, which avoids these artifacts and allows reproducible quantitative values of the diffusion coefficient to be measured in vivo. The effects of perfusion are easily visible in a phantom containing a gel. The results for human brain show a significant “perfusion fraction” in grey matter, consistent with an extracellular, possibly microvascular, volume of about 10%. © 1991 Academic Press, Inc.This publication has 15 references indexed in Scilit:
- Echo planar imaging of isolated perfused rat hearts at 4.7 T: A comparison of Langendorff and working heart preparationsMagnetic Resonance in Medicine, 1990
- Single-shot diffusion imaging at 2.0 teslaJournal of Magnetic Resonance (1969), 1990
- Minimum inductance coilsJournal of Physics E: Scientific Instruments, 1988
- Improvements in snap-shot nuclear magnetic resonance imagingThe British Journal of Radiology, 1988
- Separation of diffusion and perfusion in intravoxel incoherent motion MR imaging.Radiology, 1988
- Passive screening of switched magnetic field gradientsJournal of Physics E: Scientific Instruments, 1986
- A target field approach to optimal coil designJournal of Physics D: Applied Physics, 1986
- Active magnetic screening of gradient coils in NMR imagingJournal of Magnetic Resonance (1969), 1986
- Multi-planar image formation using NMR spin echoesJournal of Physics C: Solid State Physics, 1977
- Spin Diffusion Measurements: Spin Echoes in the Presence of a Time-Dependent Field GradientThe Journal of Chemical Physics, 1965