Evaluating doppler devices using a moving string test target
- 1 January 1982
- journal article
- research article
- Published by Wiley in Journal of Clinical Ultrasound
- Vol. 10 (1), 25-30
- https://doi.org/10.1002/jcu.1870100106
Abstract
Two moving string test targets have been developed and used to characterize ultrasound Doppler instrumentation. The characteristics investigated were sample volume size and location. The tests were performed on a combined echo/Doppler instrument and on an annular array system. The procedures can be carried out routinely in the clinical laboratory to ensure that the instrument is working properly or as an aid for correct interpretation of acquired data.Keywords
This publication has 13 references indexed in Scilit:
- Transcutaneous Blood Flow Measurements Using Pseudorandom Noise Doppler SystemIEEE Transactions on Biomedical Engineering, 1980
- Doppler methods for quantitative measurement and localization of peripheral arterial occlusive disease by analysis of the blood flow velocity waveformJapanese Journal of Clinical Oncology, 1978
- Methods of measuring the performance of ultrasonic pulse-echo diagnostic equipmentJapanese Journal of Clinical Oncology, 1977
- Doppler echocardiography. Use of a graphical display system.Circulation, 1977
- A fundamental interpretation of ultrasonic Doppler velocimetersJapanese Journal of Clinical Oncology, 1976
- Accuracy and limitations of the ultrasonic Doppler blood velocimeter and zero crossing detectorJapanese Journal of Clinical Oncology, 1975
- Ultrasonic Duplex Echo-Doppler ScannerIEEE Transactions on Biomedical Engineering, 1974
- The resolution of the ultrasound pulsed Doppler for blood velocity measurementsJournal of Biomechanics, 1973
- Technique for studying the sample volume of ultrasonic Doppler devicesMedical & Biological Engineering & Computing, 1973
- Spectral Broadening of Acoustic Reverberation in Doppler-Shift Fluid FlowmetersThe Journal of the Acoustical Society of America, 1964