European Concerted Action on Anticoagulation (ECAA) - The Multicentre Calibration of Rabbit and Human ECAA Reference Thromboplastins
- 1 December 1996
- journal article
- research article
- Published by Georg Thieme Verlag KG in Thrombosis and Haemostasis
- Vol. 76 (06), 0977-0982
- https://doi.org/10.1055/s-0038-1650695
Abstract
Two candidate ECAA reference preparations have been calibrated in a multicentre exercise at 14 representative national laboratories to provide reference thromboplastin for a large field study in 16 European countries. Two preparations were required because of the established differences in ISI arising from the two main routes of calibration via rabbit and human IRP. To be comparable with working reagents in everyday use a human plain recombinant reagent of low ISI and a rabbit plain preparation of moderately high ISI were selected. A precise calibration of the two candidate preparations has been achieved with the manual PT technique and an ISI of 0.95 (SE 0.0078) for the human reagent and 1.67 (SE 0.0322) for the rabbit reagent.Keywords
This publication has 4 references indexed in Scilit:
- The Value of Plasma Calibrants in Correcting Coagulometer Effects on International Normalized Ratios:An International Multicenter StudyAmerican Journal of Clinical Pathology, 1995
- A Proposed Scheme for Calibration of International Reference Preparations of Thromboplastin for the Prothrombin TimeThrombosis and Haemostasis, 1995
- Comparison of Two Automated Coagulometers and the Manual Tilt-Tube Method for the Determination of Prothrombin TimeAmerican Journal of Clinical Pathology, 1989
- Effect of automation on prothrombin time test in NEQAS surveys.Journal of Clinical Pathology, 1989