EVALUATION AND COMPARISON OF 2 FULLY AUTOMATED RADIOASSAY SYSTEMS WITH DISTINCTLY DIFFERENT MODES OF ANALYSIS
- 1 January 1980
- journal article
- research article
- Vol. 21 (12), 1162-1168
Abstract
Two fully automated radioimmunoassay systems with batch and sequential modes of analysis were used to assay [human] serum thyroxine, triiodothyronine and digoxin. The results obtained were compared with those obtained by manual methods. The batch system uses antibody coated tubes while the sequential system uses immobilized antibody chambers for the separation of bound from free ligands. In accuracy, both systems compared favorably with the established manual methods, but the sequential system showed better precision than the batch system. There was a statistically significant carryover of thyroxine in the sequential system when there were at least 6-fold differences in the concentrations of thyroxine in adjacent samples, but the carryover was not significant in the batch system. Compared with the batch system, the sequential system has a shorter throughtime for individual samples (time from aspiration of the sample to the printout of results), but a longer interval for final overall printout of assay results (lower throughput).This publication has 1 reference indexed in Scilit:
- Design of quality-control specimens for use with a small multi-channel analyzer.Clinical Chemistry, 1977