Quantification of mitogen induced human lymphocyte proliferation: Comparison of alamarbluetm assay to 3h‐thymidine incorporation assay

Abstract
Proliferation of human lymphocytes in response to various stimuli has traditionally been assessed by measuring uptake of radiolabeled nucleotides such as 3H‐thymidine. We have evaluated a fluorometric assay, which uses the commercially available reagent, alamarBlue, as a potential substitute for the 3H‐thymidine assay in measuring proliferation of human lymphocytes. In this assay, alamarBlueTM is added to a population of cells where it is reduced by mitochondrial enzyme activity. The reduced form of the reagent is fluorescent and can be quantitatively detected. The safety and convenience of the alamarBlueTM assay make it very attractive for use in the clinical laboratory. In this study peripheral blood mononuclear cells (PBMC) from healthy donors were stimulated using the mitogen Concanavalin A, and proliferation was assessed using either the 3H‐thymidine or the alamarBlueTM assay. The alamarBlueTM assay reliably detects human PBMC and we found that the linear range of detection was 104 cells/well (96‐well plate) to 5 × 105 cells/well. Detection of human PBMC is highly reproducible and the alamarBlue assay may be suitable in a number of applications where detection or relative quantitation of human PBMC is required. The alamarBlue assay also detected mitogen induced proliferation of PBMC although with a significantly lower level of sensitivity than the standard 3H‐thymidine assay.©1995 wiley‐Liss, inc.