An Implantable Seal‐less Centrifugal Pump with Integrated Double‐Disk Motor
- 1 July 1995
- journal article
- Published by Wiley in Artificial Organs
- Vol. 19 (7), 639-643
- https://doi.org/10.1111/j.1525-1594.1995.tb02395.x
Abstract
Thrombus formation and sealing problems at the shaft as well as the compact and efficient design of the driving unit have been major difficulties in the construction of a long-term implantable centrifugal pump. To eliminate the problems of the seal, motor size, and efficiency, two major steps were taken by modifying the Vienna implantable centrifugal pump. First, a special driving unit was developed, in which the permanent magnets of the motor themselves are used for coupling the force into the rotor. Second, the rotor shaft in the pumping chamber was eliminated by adopting a concept recently presented by Ohara. The rotor is supported by 3 pins, which run on a carbon disk, whose concave shape leads to stabilization. The device has the following specifications: size: 65 mm (diameter) by 35 mm (height), 101 cm3; priming volume 30 cm3, 240 g; and a 6-pole brushless double disk DC motor. The required input power of the described prototype is 15 W at 150 mm Hg, 5 L/min (overall eta = 11%), and has an in vitro index of hemolysis (IH) of 0.0046 g/100 L. The test for in vitro thrombus growth exhibited far less thrombus formation in the new design than in designs with axles. In conclusion, the design of a special driving unit and the elimination of the axle led to the construction of a small pump with very low blood traumatization.Keywords
This publication has 13 references indexed in Scilit:
- Recent Studies of the Centrifugal Blood Pump with a Magnetically Suspended ImpellerArtificial Organs, 1995
- In Vitro Analysis of the Floating Mode Phenomenon in the Free Impeller Gyro PumpArtificial Organs, 1995
- A Flow Visualization Study of the NCVC Centrifugal Blood PumpArtificial Organs, 1994
- The Vienna Implantable Centrifugal Blood PumpArtificial Organs, 1994
- Development of a Miniature Intraventricular Axial Flow Blood PumpAsaio Journal, 1993
- DARSTELLUNG DER STRÖMUNGSVERHÄLTNISSE IN ZENTRIFUGAL-BLUTPUMPEN MIT STANDARD-VHS-VIDEO SYSTEMENBiomedizinische Technik/Biomedical Engineering, 1993
- Effect of Stationary Guiding Vanes on Improvement of the Washout Behind the Rotor in Centrifugal Blood PumpsAsaio Journal, 1992
- Motor design with large air gap for centrifugal blood pumps using rare-earth magnetsElectrical Engineering, 1990