A Longitudinal Study of the Age-Adjusted Variability of Automated Visual Fields
- 1 August 1987
- journal article
- research article
- Published by American Medical Association (AMA) in Archives of Ophthalmology (1950)
- Vol. 105 (8), 1083-1086
- https://doi.org/10.1001/archopht.1987.01060080085033
Abstract
• Twenty-six normal eyes of 26 subjects were tested three times during an 18-month period using the 30-2 program of the Humphrey Visual Field Analyzer (Humphrey Instruments, San Leandro, Calif). For the average eye, the SD of threshold values over time ranged from 1.8 to 5.5 dB, depending on the location in the field. For those subjects under 60 years of age, SDs ranged from 1.2 to 2.8 dB. For those subjects 60 years of age and older, the range was 2.2 to 8.3 dB. Short-term fluctuations contributed 25% of the variability of threshold measurements over time. Short-term and long-term variations were greater in older eyes, but long-term variability was more affected by age than was short-term variation.This publication has 7 references indexed in Scilit:
- Age-Related Changes of the Normal Visual FieldArchives of Ophthalmology (1950), 1986
- Influence of Age on the Visual Fields of Normal SubjectsAmerican Journal of Ophthalmology, 1986
- Asymmetry and Variation in the Normal Hill of VisionArchives of Ophthalmology (1950), 1986
- Evaluation of Nerve Fiber Layer AssessmentArchives of Ophthalmology (1950), 1984
- Variation in Visual Field Measurements with an Automated PerimeterAmerican Journal of Ophthalmology, 1984
- The Estimation and Testing of the Components of Long-Term Fluctuation of the Differential Light ThresholdDocumenta Ophthalmologica Proceedings Series, 1983
- STATIC PERIMETRY: ACCURACY AND FLUCTUATIONSActa Ophthalmologica, 1976