'Minified' Goldmann Applanating Prism for Tonometry in Monkeys and Humans
- 1 March 1980
- journal article
- research article
- Published by American Medical Association (AMA) in Archives of Ophthalmology (1950)
- Vol. 98 (3), 542-546
- https://doi.org/10.1001/archopht.1980.01020030538022
Abstract
• The tip diameter of the standard Goldmann applanating prism was reduced from 7.0 mm to 4.0 or 4.5 mm, but the endpoint of intraocular pressure measurement—applanation of a circle of cornea having a diameter of 3.06 mm—was not changed. The "minified" tonometers exhibited standard mechanical calibration characteristics in three standard configurations. The 4.5-mm tonometer was calibrated for cynomolgus monkey eyes in vivo by open and closed stopcock manometry. By open stopcock manometry, between pressures of 4 and 70 mm Hg, IOP(tonometer) = 1.01 IOP(manometer) – 0.72, with very little scatter or curvature. During closed stopcock manometry, between pressures of 5 and 55 mm Hg, IOP(tonometer) = 1.07 IOP(manometer)– 1.32, again with very little scatter or curvature. The tonometer elevated manometric IOP by an average of about 7.5%; the percentage pressure elevation decreased slightly as pressure increased. The 4.0-mm tonometer was calibrated against a standard tonometer in living human eyes. Over the pressure range from 0 to 75 mm Hg, IOP(minitonometer) = 0.98 IOP(standard tonometer)+ 0.82, with very little scatter or curvature. Similar comparison of two standard tonometers showed a virtually identical relationship and scatter. We conclude that minified Goldmann tonometers provide accurate and reproducible measurements of IOP in cynomolgus monkey and human eyes.This publication has 10 references indexed in Scilit:
- The Goldmann Applanation TonometerClinical and Experimental Optometry, 1976
- APPLANATION TONOMETRY IN THE CONSCIOUS CYNOMOLGUS MONKEY (MACACA FASCICULARIS)Acta Ophthalmologica, 1976
- EFFECTS OF REPEATED TONOMETRY: GENUINE AND SHAM MEASUREMENTSActa Ophthalmologica, 1972
- THE MACKAY‐MARG TONOMETERActa Ophthalmologica, 1971
- Intraocular Pressure Following Penetrating KeratoplastyAmerican Journal of Ophthalmology, 1969
- Repeated Applanation TonometryAmerican Journal of Ophthalmology, 1968
- Electronic Applanation TonometryAmerican Journal of Ophthalmology, 1966
- Simple Hand Applanation Tonometer: For use on the Seated as Well as on the Supine PatientAmerican Journal of Ophthalmology, 1966
- A Clinical Evaluation of the Electronic Applanation TonometerAmerican Journal of Ophthalmology, 1966
- Hand-held applanation tonometer.British Journal of Ophthalmology, 1965