Toward Real-World Sequencing by Microdevice Electrophoresis
Open Access
- 1 September 1999
- journal article
- Published by Cold Spring Harbor Laboratory in Genome Research
- Vol. 9 (9), 853-858
- https://doi.org/10.1101/gr.9.9.853
Abstract
We report results using a microdevice for DNA sequencing using samples from chromosome 17, obtained from the Whitehead Institute Center for Genome Research (WICGR) production line. The device had an effective separation distance of 11.5 cm and a lithographically defined injection width of 150 μm. The four-color raw data were processed, base-called by the sequencing software Trout, and compared to the corresponding ABI 377 sequence from WICGR. With a criteria of 99% accuracy, we achieved average continuous reads of 505 bases in 27 min with 3% linear polyacrylamide (LPA) at 150 V/cm, and 460 bases in 22 min with 4% LPA at 200 V/cm at a temperature of 45°C. In the best case, up to 565 bases could be base-called with the same accuracy in R as low as R = 0.35. This may be due in part to the high signal-to-noise ratio of the microdevice. Unlike many results reported on capillary machines, no additional sample cleanup other than ethanol precipitation was required. In addition, DNA fragment biasing (i.e., discrimination against larger fragments) was reduced significantly through the unique sample injection mechanism of the microfabricated device. This led to increased signal strength for long fragments, which is of great importance for the high performance of the microdevice.Keywords
This publication has 19 references indexed in Scilit:
- Two-Color Multiplexed Analysis of Eight Short Tandem Repeat Loci with an Electrophoretic MicrodeviceAnalytical Biochemistry, 1999
- Characterization of high molecular mass linear polyacrylamide powder prepared by emulsion polymerization as a replaceable polymer matrix for DNA sequencing by capillary electrophoresisElectrophoresis, 1998
- Rapid DNA Sequencing of More Than 1000 Bases per Run by Capillary Electrophoresis Using Replaceable Linear Polyacrylamide SolutionsAnalytical Chemistry, 1996
- Microchip Electrophoretic Immunoassay for Serum CortisolAnalytical Chemistry, 1996
- Ultra-High-Speed DNA Sequencing Using Capillary Electrophoresis ChipsAnalytical Chemistry, 1995
- Separation of Fragments up to 570 Bases in Length by Use of 6% T Non-Cross-Linked Polyacrylamide for DNA Sequencing in Capillary ElectrophoresisAnalytical Chemistry, 1994
- Micromachining a Miniaturized Capillary Electrophoresis-Based Chemical Analysis System on a ChipScience, 1993
- Analysis of resolution in DNA sequencing by capillary gel electrophoresisThe Journal of Physical Chemistry, 1993
- Bias in quantitative capillary zone electrophoresis caused by electrokinetic sample injectionAnalytical Chemistry, 1988
- High-performance electrophoresisJournal of Chromatography A, 1985