Presented here is a theoretical analysis of the recently developed thermal pulse decay (TPD) method for a simultaneous measurement of local tissue conductivity and blood perfusion rate. The paper describes the theoretical model upon which the TPD method is based and details its capabilities and limitations. The theoretical aspects that affected the development of the measurement protocol are also discussed. The performance of the method is demonstrated with an experimental example which compares the measurements of local kidney blood perfusion rates made using the TPD method with the total renal blood flow obtained coincidentally using a blood flowmeter, in an anesthetized dog.Keywords
This publication has 5 references indexed in Scilit:
- Thermal conductivity and H2O content in rabbit kidney cortex and medullaJournal of Thermal Biology, 1983
- Measurement of glomerular blood flow in rabbits and rats: Erroneous findings with 15-µm microspheresKidney International, 1979
- Thermal quantification of region myocardial perfusion and heat generationAmerican Journal of Physiology-Heart and Circulatory Physiology, 1979
- Evaluation of methods of measuring glomerular and nutrient blood flow in rat kidneysAmerican Journal of Physiology-Heart and Circulatory Physiology, 1978
- STUDIES OF THERMAL INJURY .2. THE RELATIVE IMPORTANCE OF TIME AND SURFACE TEMPERATURE IN THE CAUSATION OF CUTANEOUS BURNS1947