Leukocyte depletion filters: A comparison of efficiency
- 1 November 1990
- journal article
- research article
- Published by Wiley in American Journal of Hematology
- Vol. 35 (3), 184-186
- https://doi.org/10.1002/ajh.2830350308
Abstract
Four commercially available leukocyte depletion filters were evaluated using a flow cytometric technique to determine the efficiency of each type of filter. The Sepacell R-500A® and the PALL RC50® were the most efficient in leukocyte depletion having a mean depletion percentage of 99.3% and 99.5%, respectively. The PALL RC100® which is used for two units also had a 99.3% depletion with the first unit but with the second unit the depletion dropped to 94.2%. The Imugard IG-500® had a 97.3% depletion. Red cell recovery ranged from 87.4% for the PALL RC50® to 92.2% for the Sepacell R-500A®.Keywords